Disposable temperature measuring line

http://120.24.179.97:8069/web/image/product.template/3711/image_1920?unique=dbcb708

Scope of application

Thermometers, bare probes, are mainly used as temperature measurement components for disposable urinary catheters, but also for other products that require accurate temperature measurement.
Compatible with a wide range of monitors, providing accurate body temperature signals.
Single-use to avoid the risk of cross-contamination.

characteristic

Main performance: It is used to collect and transmit physiological signals of patients' body temperature.
Scope of application: Suitable for different models of multi-parameter monitors and electronic thermometers that can be continuously measured.
Temperature measurement range: 32 ºC~ 44 ºC, maximum allowable error: ±0.1 ºC
Response time: <150s
Classification according to the degree of electric shock: According to the general electric shock protection classification rules, the thermometer belongs to the BF type, and the application part of the body temperature probe is also the BF type after being matched with it.
Classified according to the degree of protection against liquid ingress: ordinary equipment.
Classification by mode of operation: continuous operation.
Classified according to the degree of safety when used in the case of flammable anesthetic gas mixed with air or flammable anesthetic gas mixed with oxygen or nitrous oxide: non-AP/APG.
Electromagnetic compatibility: in line with the requirements of YY 0505-2012.
Probe (resistive) maximum energy dissipation: When immersed in a tank at a temperature of 37±0.1°C, the maximum power provided of 0.06mW should not result in an energy dissipation that raises the temperature by more than 0.02°C.
Stability: After the temperature probe is placed in an environment of (80±2)°C for 96 hours, the maximum allowable error value before and after the probe exposure meets the requirements of ±0.1 ºC.
Body Fluid Protection: The body temperature probe is insulated enough to prevent changes in the indicated temperature from exceeding ±0.02°C when immersed in conductive fluids.
Cleaning and disinfection: After the temperature probe is cleaned and disinfected according to the cleaning and disinfection method recommended in the manual, the maximum allowable error meets the requirements of ±0.1 ºC.